Skip to main content

Cytostatic Biosafety Cabinets

Cyto-2 and Cyto-3

The Cruma Cyto cabinets, Class II A 2, have been especially designed for the preparation of cytostatic medicine, as well as for handling biological agents included in risk groups 1, 2 and 3. Their operating scheme includes an additional third filtration stage under the work surface. What’s more, the patented Bag-In Bag-Out system for filter changes means the contaminated areas remain completely airtight as far as the outside is concerned when the third filtration stage which is located under the work surface is being replaced. Thus the maintenance technician has absolutely no physical contact with the contaminated filter or with the areas that have a high risk of potential contamination.

Features

Yes, they all fit.

  • EN-12469 and DIN 12980 certification by TUV NORD
  • Three stages of H14 class High Efficiency Particulate Air filters with 99.995% efficiency for .3 micron particles (most penetrating particle diameter) (EN 1822-1 and EN 13091:1999 tested and certified)
  • Tertiary filter change through the “bag in- bag out technology”. Avoids physical contact with the third stage of filtration during filter replacement.
  • Sliding sash window with smart control
  • State of the art Microprocessor control system offering:

– High resolution digital screen
– Automatic control of preset airflow volumes
– Sliding sash window with smart control
– Permanent monitoring of HEPA filters life span
– Permanent display of working conditions
– Maintains air flow stability in the case of progressive filter clogging
– Low barrier alarm
– Power failure alarm

  • Multilayer 6mm safety glass
  • Comfortable 200 mm front opening
  • Easy to install retrofit options through lateral sides
  • Sloped back side of the working chamber for the best down flow distribution
  • Front barrier air speed ≥0.5mt/sec
  • Light intensity on work surface >1200 lux
  • Noise level < 55 dB
  • Easily installed exhaust duct (optional)
  • Safety key to avoid unwanted operation
  • Self calibration cycle performed when cabinet is switched on
  • Interconnected UV and fluorescent lights
  • Stainless steel worktop with 2B finish (including spillage tray)
  • In case of power failure, the cabinet is re-set to original working conditions

Handling

Biological Hazard Works

  • Preparation of cytostatic drugs
  • Manipulation of biological agents of risk group 1, 2 and 3.

I want more information and Prices

    Accept the privacy policy





    *Required fields

    Models and Technical Specifications

    Two sizes for all needs.

    Model Cyto-2 Cyto-3
    Internal Sizes (Width/Depth/High) 1230x600x700 1830x600x700
    External Sizes (Width/Depth/High) 1380x840x2220 1990x840x2220
    Filtration columns 2 2
    Filter number 3 3
    Fans 2 2
    Average volume of treated air/hour +/- 400 m3/h +/-600 m3/h
    Average face velocity 0.50 m/s 0.50 m/s
    Total electrical power consumption 360 W 650 W
    Voltage-Frequency 110-220 V / 50-60 Hz 110-220 V / 50-60 Hz
    Light intensity Fluorescent 36W / 1200Lux Fluorescent 36W / 1200Lux
    UV Lamp Optional Optional
    Noise level 55 dB 57 dB
    Weight 260 Kg cabinet – 100 Kg base 390 Kg cabinet – 120 Kg base
    Packaging volume 2.42 m3 cabinet – 1.63 m3 base 3.45 m3 cabinet – 2.32 m3 base
    OFFER REQUEST I WANT MORE INFO & PRICES I WANT MORE INFO & PRICES

    I want more information and Prices

      Accept the privacy policy





      *Required fields

      Cruma HEPA Filters

      Two + One. For Pure Air and Total Safety.

      Double-filter plus a third additional stage under the work surface. The patented Bag-In Bag-Out system for filter changes means the contaminated areas remain completely airtight as far as the outside is concerned when the third filtration stage which is located under the work surface is being replaced.

      Main H14 HEPA filter with a flow equalizer and an H14 safety filter at the point of discharge, both of which have an efficiency rate of over 99.995 % for particles of 0.3 microns (Standards MIL-F51068C, EN-ISO 14644-1 and BS 3298).EN-ISO 14644-1 y BS 3298).

      * To view the technical sheets of our filters, please contact Cruma sales department at export@cruma.es
      – It is the user’s responsibility to correctly manage the waste generated from filters based on the use given to them and according to the applicable requirements.
      – The supply of spare parts is guaranteed during a period of 10 years

      I want more information and prices

        Accept the privacy policy





        *Required fields

        Any doubt about what to choose?  Take our fantastic test!